Installation And Use Of Coriolis Mass Flowmeter

In order for the Coriolis mass flowmeter to work normally, safely and with high performance, it is very important to install and use it correctly. This includes mechanical installation, electrical installation and maintenance.

(1) Installation
The following issues should be noted during mechanical installation:

a. The flow sensor should be installed on a solid foundation. Small-diameter mass flowmeters with an inner diameter of less than 10mm are installed on a balanced, hard and vibration-free bottom surface, such as a wall, ground or a special foundation. If used in a high-vibration environment, attention should be paid to the vibration absorption of the foundation, and flexible pipes should be used to connect the sensor inlet and outlet to the pipeline; larger-diameter flowmeters are directly installed on the process pipeline, and the flowmeter should be firmly fixed with pipe clamps and supports.

b. To prevent mutual influence between CMFs, when multiple flowmeters are used in series or parallel, the distance between each flow sensor should be far enough, and the pipe clamps and supports should be set on their own independent foundations.

C. To ensure that there is no accumulation of gas or liquid residual value in the flow sensor during use, for elbow-type flowmeters, the elbow should be downward when measuring liquids and upward when measuring gases. When measuring slurry or discharge liquid, the sensor should be installed in a vertical pipe with the flow direction from bottom to top. For straight tube flowmeters, avoid installing it at the highest point when installing horizontally to prevent air accumulation. When connecting the sensor and the process pipeline, it must be installed without stress, especially for flow sensors with some straight tube measuring tubes. (2) Use and maintenance ① Flowmeter zero point adjustment. After the flow sensor is filled with the measured fluid, close the downstream valve of the sensor and adjust the zero point of the flowmeter under conditions close to the working temperature. When adjusting the zero point, it is very important to ensure that the downstream valve is completely closed and confirm that there is no fluid leakage. If there is leakage in the valve during zero adjustment, it will cause a large error in the entire measurement. ② Set the flow and density calibration coefficient. Correctly setting the flow and density calibration coefficients is very important for the work of the flowmeter. The flow calibration coefficient represents the sensitivity of the sensor and the flow temperature coefficient. The sensitivity indicates how much flow is measured per microsecond time difference. (The unit is usually g/s); the flow temperature coefficient indicates the degree to which the elastic modulus of the sensor is affected by temperature. These factors are directly related to the flowmeter's measurement accuracy. The density calibration factor represents the sensor's natural oscillation period (usually expressed in microseconds) and the density temperature coefficient at 0°C when the tube contains air and water. Clearly, these factors are directly related to the accuracy of density measurements.
③ Maintaining the proper functioning of the flowmeter during use. Promptly detecting and troubleshooting faults is crucial for proper flowmeter operation. Common fault conditions in practice include the following:
a. No output: This refers to flow passing through the sensor but no signal output;
b. No output change: The output remains unchanged despite changes in flow rate;
c. Abnormal output: The output fluctuates randomly, unrelated to changes in flow rate;
d. Intermittent output: This refers to intermittent output with irregular starts and stops, but when present, the output signal accurately reflects the flow rate.

The above fault conditions should be carefully inspected and corrected, and the manufacturer should be consulted for repairs if necessary.